METER BARO Module

BARO INTEGRATOR GIYA
DESKRIPSIYON SA SENSOR
Ang BARO Module usa ka tukma nga barometer aron mabayran ang mga potensyal nga pagsukod sa matric sa TEROS 31 ug TEROS 32 nga mga tensiometer. Ang BARO Module mahimong magamit isip usa ka standalone sensor aron mabayran ang usa o daghan pa nga mga tensometer sa usa ka lugar nga pagsukod, o ingon usa ka digital/analog converter aron mabayran ang usa ka konektado nga TEROS 31 o TEROS 32 nga kantidad ug i-convert ang SDI-12 signal sa usa ka analog vol.tage output (8-pin nga bersyon lamang). Ang BARO Module ug TEROS 32 nga kombinasyon mahimong gamiton isip T8 tensiometer replacement. Para sa mas detalyado nga deskripsyon kung giunsa paghimo niini nga sensor ang pagsukod, tan-awa ang BARO Module User Manual.

MGA APLIKASYON
- Barometric pressure measurement
- Barometric compensation of matric potential measurements
- Digital/analog converter for directly connected TEROS 31 and TEROS 32 tensiometers
- Appropriate for non-METER data loggers to connect TEROS 31 and TEROS 32
ADVANTAGES
- Ang digital sensor nagpahibalo sa daghang mga pagsukod sa usa ka serial interface
- Ubos nga input voltage kinahanglanon
- Ang disenyo sa ubos nga gahum nagsuporta sa mga data logger nga gipadagan sa baterya
- SDI-12, Modbus RTU or tensio LINK serial communications protocol supported
- Analog output supported (only 8-pin version)
ESPESPIKASYON
| MGA ESPESPIKASYON SA PAGSUKO | |
| Barometric Pressure | |
| Sakop | + 65 kPa hangtod sa +105 kPa |
| Resolusyon | ± 0.0012 kPa |
| Pagkatukma | ± 0.05kPa |
| Temperatura | |
| Sakop | -30 ngadto sa + 60 °C |
| Resolusyon | ± 0.01 °C |
| Pagkatukma | ± 0.5 °C |
| MGA ESPESPIKASYON SA KOMUNIKASYON | |
| Output | |
| Analog Output (8-pin connector only)0 to 2,000 mV (default)0 to 1,000 mV (configurable with tensio VIEW) | |
| Digital OutputSDI-12 communications protocol tensio LINK communication protocol Modbus RTU communication protocol | |
| Pagkaangay sa Data Logger | |
| Analog Output Bisan unsang data acquisition system nga makahimo sa pagbalhin sa 3.6- to 28-VDC excitation ug single-ended o differential voltage pagsukod sa mas dako pa sa o katumbas sa 12-bit nga resolusyon. | |
| Digital Output Bisan unsang data acquisition system nga makahimo sa 3.6- to 28-VDC excitation ug RS-485 Modbus o SDI-12 nga komunikasyon. | |
| PISIKAL NGA ESPESPIKASYON | |
| Mga sukat | |
| Gitas-on | 80 mm (3.15 in) |
| Lapad | 29 mm (1.14 in) |
| Taas | 30 mm (1.18 in) |
| Gitas-on sa Cable | |
| 1.5 m (standard)NOTE: Contact Customer Support if a nonstandard cable length is needed. | |
| Mga Uri sa Konektor | |
| 4-pin and 8-pin M12 plug connector or stripped and tinned wires | |
| PAGSUNOD | |
| EM ISO/IEC 17050:2010 (CE Mark) | |
KATAPUSANG SIRCUIT UG MGA TYPE SA KONEKSIYON
Refer to Figure 2 to connect the BARO Module to a data logger. Figure 2 provides a low-impedance variant of the recommended SDI-12 specification.



BARO MODULE INTEGRATOR GIYA 



PANAGANA
Ang mga sensor sa METER gihimo sa pinakataas nga mga sumbanan, apan ang sayop nga paggamit, dili husto nga proteksyon, o dili husto nga pag-instalar mahimong makadaut sa sensor ug posibleng makawala sa warranty. Sa dili pa i-integrate ang mga sensor ngadto sa sensor network, sunda ang girekomendar nga mga instruksyon sa pag-instalar ug ipatuman ang mga panalipod aron mapanalipdan ang sensor gikan sa makadaot nga interference.
SENSOR KOMUNIKASYON
METER digital sensors feature a serial interface with shared receive and transmit signals for communicating sensor measurements on the data wire. The sensor supports SDI-12, tensio LINK, and Modbus over RS-485 two-wire. The sensor automatically detects the interface and protocol which is being used. Each protocol has implementation advantages ug mga hagit. Palihug kontaka ang METER Customer Support kung dili klaro ang pagpili sa protocol alang sa gusto nga aplikasyon.
- SDI-12 PASIUNA
Ang SDI-12 kay usa ka standards-based protocol para sa interfacing sensors sa data loggers ug data acquisition equipment. Daghang mga sensor nga adunay talagsaon nga mga adres mahimong mag-ambit sa usa ka sagad nga 3-wire bus (gahum, yuta, ug data). Ang duha ka paagi nga komunikasyon tali sa sensor ug logger posible pinaagi sa pagpaambit sa linya sa datos alang sa pagpadala ug pagdawat ingon nga gihubit sa sumbanan. Ang mga pagsukod sa sensor gi-trigger sa protocol command. Ang SDI-12 protocol nanginahanglan usa ka talagsaon nga alphanumeric sensor address alang sa matag sensor sa bus aron ang usa ka data logger makapadala mga mando ug makadawat mga pagbasa gikan sa piho nga mga sensor.
I-download ang SDI-12 Specification v1.3 aron makat-on og dugang mahitungod sa SDI-12 protocol. - RS-485 PASIUNA
Ang RS-485 usa ka lig-on nga pisikal nga koneksyon sa bus aron makonektar ang daghang mga aparato sa usa ka bus. Kini makahimo sa paggamit sa taas kaayo nga mga distansya sa cable ubos sa mapintas nga mga palibot. Imbes nga SDI-12, ang RS-485 naggamit ug duha ka dedikadong wires para sa signal sa data. Gitugotan niini ang paggamit sa mas taas nga mga kable ug mas dili sensitibo sa pagpanghilabot gikan sa gawas nga mga gigikanan, tungod kay ang signal adunay kalabotan sa lainlaing mga wire, ug ang mga sulog sa suplay dili makaimpluwensya sa signal sa datos. Tan-awa ang Wikipedia para sa dugang nga mga detalye sa RS-485. - TENSIOLINK RS-485 PASIUNA
Ang tensioLINK usa ka paspas, kasaligan, proprietary serial communications protocol nga nagkomunikar pinaagi sa RS-485 interface. Kini nga protocol gigamit sa pagbasa sa datos ug pag-configure sa mga feature sa device. Ang METER naghatag ug tensioLINK PC USB converter ug software aron direktang makigkomunikar sa sensor, pagbasa sa datos, ug pag-update sa firmware. Palihug kontaka ang Suporta sa Customer alang sa dugang nga impormasyon bahin sa tensionLINK. - MODBUS RTU RS-485 PASIUNA
Ang Modbus RTU usa ka komon nga serial communications protocol nga gigamit sa Programmable Logic Controllers (PLCs) o data loggers aron makigkomunikar sa tanang matang sa digital device. Ang komunikasyon naglihok sa pisikal nga RS-485 nga koneksyon. Ang kombinasyon sa RS-485 alang sa pisikal nga koneksyon ug Modbus isip serial communications protocol nagtugot sa paspas ug kasaligan nga pagbalhin sa datos alang sa taas nga gidaghanon sa mga sensor nga konektado sa usa ka serial bus wire. Gamita ang mosunod nga mga link para sa dugang impormasyon sa Modbus: Wikipedia ug modbus.org. - PAG-INTERFACING SA SENSOR NGADTO SA COMPUTER
The serial signals and protocols supported by the sensor require some type of interface hardware to be compatible with the serial port found on most computers (or USB-to-serial adapters). There are several
SDI-12 interface adapters available in the marketplace; however, METER has not tested any of these interfaces and cannot make a recommendation as to which adapters work with METER sensors. METER data loggers and the ZSC handheld device can operate as a computer-to-sensor interface for making on-demand sensor measurements.
Ang BARO Module mahimo usab nga i-configure ug masukod pinaagi sa tensioLINK gamit ang METER software tensionVIEW, magamit aron ma-download sa meter.ly/software. Aron makonektar ang usa ka BARO Module sa usa ka kompyuter gikinahanglan ang usa ka tensoLINK USB converter ug usa ka angay nga adapter cable. - METER SDI-12 IMPLEMENTATION
Kung ang BARO Module konektado tali sa TEROS 31 o 32 tensiometer, ang barometric air pressure ug ang absolute pressure sa TEROS tensiometer mahimong mabasa pinaagi sa Modbus. Ang kompensasyon nga potensyal sa matrix mahimong mabasa usab pinaagi sa Modbus.
METER sensors use a low-impedance variant of the SDI-12 standard sensor circuit (Figure 2). During the power-up time, sensors output some sensor diagnostic information and should not be communicated with until the power-up time has passed. After the power up time, the sensors are fully compatible with all commands listed in the SDI-12 Specification v1.3 except for the continuous measurement commands ( aR0 – aR9 and aRC0 – aRC9 ). M , R , and C command implementations are found on pages 8–9. Out of the factory, all METER sensors start with SDI-12 address 0 . - MGA KONSIDERASYON SA SENSOR BUS
SDI-12 sensor buses require regular checking, sensor upkeep, and sensor troubleshooting. If one sensor goes down, that may take down the whole bus even if the remaining sensors are functioning normally. Power cycling the SDI-12 bus when a sensor is failing is acceptable. METER SDI-12 sensors can be power-cycled and read on the desired measurement interval or powered continuously and commands sent when a measurement is desired based on specified communication timing. Many factors influence the effectiveness of the bus configuration. Visit metergroup.com alang sa mga artikulo ug virtual nga mga seminar nga adunay daghang impormasyon.
SDI-12 CONFIGURATION
Ang talaan 1 naglista sa SDI-12 nga configuration sa komunikasyon.
| Lamesa 1 Konfigurasyon sa komunikasyon sa SDI-12 | |
| Baud Rate | 1,200 |
| Pagsugod Bits | 1 |
| Mga Bit sa Data | 7 (LSB una) |
| Mga Parity Bits | 1 (ganan) |
| Hunong ang mga Bits | 1 |
| lohika | Baliktad (aktibo ubos) |
SDI-12 TIMING
All SDI-12 commands and responses must adhere to the format in Figure 9 on the data line. Both the command and response are preceded by an address and terminated by a carriage return and line feed combination ( <CR><LF> ) and follow the timing shown in Figure 10.


KOMMON SDI-12 NGA MGA SUGO
Kini nga seksyon naglakip sa mga lamesa sa komon nga SDI-12 nga mga sugo nga sagad gigamit sa usa ka SDI-12 nga sistema ug ang katugbang nga mga tubag gikan sa METER sensors.
SUGO SA PAGKILALA ( aI! )
Ang Identification command mahimong magamit aron makakuha og lain-laing mga detalyadong impormasyon bahin sa konektado nga sensor. Usa ka example sa sugo ug tubag gipakita sa Example 1, diin ang sugo kay bold ug ang tubag nagsunod sa sugo.
Example 1 1I!113METER␣ ␣ ␣BARO␣
|
Parameter |
Naayo nga Kinaiya Gitas-on | Deskripsyon |
| 1 ako! | 3 | Data logger command.Request to the sensor for information from sensor address 1 . |
| 1 | 1 | Sensor address.Prepended on all responses, this indicates which sensor on the bus is returning the following information. |
| 13 | 2 | Nagpakita nga ang target sensor nagsuporta sa SDI-12 Specification v1.3. |
| METER ␣ ␣ ␣ | 8 | Vendor identification string.( METER and three spaces ␣ ␣ ␣ for all METER sensors) |
| BARO␣ | 6 | Sensor model string.This string is specific to the sensor type. For the BARO, the string is BARO . |
| 100 | 3 | Sensor version.This number divided by 100 is the METER sensor version (e.g., 100 is version 1.00). |
| BARO-00001 | ≤13,variable | Sensor serial number. Kini usa ka variable nga gitas-on nga field. Mahimong dili kini iapil alang sa mga tigulang nga sensor. |
BAG-O ANG ADDRESS COMMAND ( aAB! )
The Change Address command is used to change the sensor address to a new address. All other commands support the wildcard character as the target sensor address except for this command. All METER sensors have a default address of 0 (zero) out of the factory. Supported addresses are alphanumeric (i.e., A – Z , and 0 – 9 ). An exampAng output gikan sa METER sensor gipakita sa Example 2, diin ang sugo kay bold ug ang tubag nagsunod sa sugo.
Exampug 2 1A0!0
|
Parameter |
Naayo nga Kinaiya Gitas-on | Deskripsyon |
| 1A0! | 4 | Data logger command. Request to the sensor to change its address from 1 to a new address of 0 . |
| 0 | 1 | Bag-ong sensor address. Alang sa tanang sunod nga mga sugo, kining bag-ong adres gamiton sa target sensor. |
PAGPATUMAN SA COMMAND
Ang mosunod nga mga lamesa naglista sa mga may kalabutan nga Pagsukod ( M ), Padayon ( R ), ug Kadungan ( C ) nga mga sugo ug sunod nga Data ( D ) nga mga sugo, kon gikinahanglan.
MGA SUGO SA PAGSUKO SA IMPLEMENTASYON
Measurement ( M ) commands are sent to a single sensor on the SDI-12 bus and require that subsequent Data ( D ) commands are sent to that sensor to retrieve the sensor output data before initiating communication with another sensor on the bus. Please refer to Table 2 and for an explanation of the command sequence and to Table 5 for an explanation of response parameters.
Talaan 2 sa buntag! han-ay sa sugo
| Sugo | tubag |
| Kini nga sugo nagreport sa kasagaran, natipon, o pinakataas nga mga kantidad. | |
| aM! | atttn |
| aD0! | a± ± + |
| Mga komento | When a slave TEROS tensiometer is connected, <Press> hold the barometric compensated tensiometer output. If the BARO module is used in standalone <Press> returns the current barometric pressure. |
| NOTE: The measurement and corresponding data commands are intended to be used back to back. After a measurement command is processed by the sensor, a service request a <CR><LF> is sent from the sensor signaling the measurement is ready. Either wait until seconds have passed or wait until the service request is received before sending the data commands. See the SDI-12 Specifications v1.3 | |
PAHINUMDOM: Ang pagsukod ug katugbang nga mga sugo sa datos gituyo nga gamiton balik-balik. Human maproseso sa sensor ang sugo sa pagsukod, usa ka hangyo sa serbisyo a gipadala gikan sa sensor nga nagsenyas nga andam na ang pagsukod. Maghulat hangtod molabay ang ttt segundo o maghulat hangtod madawat ang hangyo sa serbisyo sa dili pa ipadala ang mga mando sa datos. Tan-awa ang SDI-12 Specifications v1.3 nga dokumento para sa dugang nga impormasyon.
KONCURRENT MEASUREMENT COMMANDS IMPLEMENTATION
Ang mga mando sa Concurrent Measurement (C) kasagarang gigamit sa mga sensor nga konektado sa usa ka bus. Ang mga sugo sa C alang niini nga sensor nagtipas gikan sa standard nga pagpatuman sa C command. Una, ipadala ang C command, hulata ang espesipikong gidugayon sa oras nga detalyado sa tubag sa C command, ug dayon gamita ang D commands aron basahon ang tubag niini sa dili pa makigkomunikar sa laing sensor.
Please refer to Table 3 for an explanation of the command sequence and to Table 5 for an explanation of response parameters.
| Table 3 aC! measurement command sequence | |
| Sugo | tubag |
| Kini nga mando nagreport dayon sa mga kantidad. | |
| aC! | atttnn |
| aD0! | a± ± + |
| NOTE: The measurement and corresponding data commands are intended to be used back to back. After a measurement commanc is pro- cessed by the sensor, a service request a<CR><LF> is sent from the sensor signaling the measurement is ready. Either wait until ttt sec- onds have passed or wait until the service request is received before sending the data commands. Please see the SDI-12 Specifications v1.3 document for more information. | |
NOTE: The measurement and corresponding data commands are intended to be used back to back. After a measurement commanc is pro-cessed by the sensor, a service request a<CR><LF> is sent from the sensor signaling the measurement is ready. Either wait until ttt sec-onds have passed or wait until the service request is received before sending the data commands. Please see the SDI-12 Specifications v1.3 document for more information.
PADAYON NGA PAGSUKOD MGA SUGO IMPLEMENTASYON
Continuous Measurement ( R ) commands trigger a sensor measurement and return the data automatically after the readings are completed without needing to send a D command. aR0! returns more characters in its response than the 75-character limitation called out in the SDI-12 Specification v1.3. It is recommended to use a buffer that can store at least 116 characters.
Please refer to Table 4 for an explanation of the command sequence and see Table 5 for an explanation of response parameters.
| Table 4 aR0! measurement command sequence | |
| Sugo | tubag |
| Kini nga sugo nagreport sa kasagaran, natipon, o pinakataas nga mga kantidad. | |
| aR0! | a± ± + |
| PAHINUMDOM: Kini nga sugo wala magsunod sa SDI-12 response timing. Tan-awa ang METER SDI-12 Implementation para sa dugang nga impormasyon. | |
NOTE: This command does not adhere to the SDI-12 response timing. See METER SDI-12 Implementation for more information.
MGA PARAMETER
Table 5 lists the parameters, unit measurement, and a description of the parameters returned in command responses for the BARO Module.
| Lamesa 5 Mga Deskripsyon sa Parameter | ||
| Parameter | Unit | Deskripsyon |
| ± | — | Positibo o negatibo nga timaan nga nagpaila sa timaan sa sunod nga kantidad |
| a | — | SDI-12 nga adres |
| n | — | Gidaghanon sa mga sukod (fixed nga gilapdon sa 1) |
| nn | — | Gidaghanon sa mga sukod nga adunay nag-unang zero kung gikinahanglan (fixed nga gilapdon sa 2) |
| ttt | s | Ang labing kadaghan nga pagsukod sa oras ang makuha (naayo nga gilapdon sa 3) |
| — | Tab nga karakter | |
| — | Ang karakter sa pagbalik sa karwahe | |
| — | Linya nga feed karakter | |
| — | ASCII character denoting the sensor type For BARO Module, the character is ; | |
| — | METER serial checksum | |
| — | METER 6-bit nga CRC |
METER MODBUS RTU SERIAL IMPLEMENTATION
Modbus over Serial Line is specified in two versions – ASCII and RTU. BARO Modules communicate using RTU mode exclusively. The following explanation is always related to RTU. Table 6 lists the Modbus RTU communication and configuration.
| Lamesa 6 Modbus communication characters | |
| Baud Rate (bps) | 9,600 bps |
| Pagsugod Bits | 1 |
| Mga Bit sa Data | 8 (LSB una) |
| Mga Parity Bits | 0 (wala) |
| Hunong ang mga Bits | 1 |
| lohika | Standard (aktibo nga taas) |
Figure 11 shows a message in RTU format. The size of the data determines the length of the message. The format of each byte in the message has 10 bits, including Start and Stop Bit. Each byte is sent from left to right: Least Significant Bit (LSB) to Most Significant Bit (MBS). If no parity is implemented, an additional stop bit is transmitted to fill out the character frame to a full 11-bit asynchronous character.
The Modbus application layer implements a set of standard Function codes divided into three categories: Public, User-defined, and Reserved. Well-defined public function codes for BARO Modules are documented in the Modbus Organization, Inc. (modbus.org) community.
For a reliable interaction between the BARO Module and a Modbus Master, a minimum 50ms delay is required between every Modbus command sent on the RS-485 bus. An additional timeout is needed for every Modbus query; this timeout is device-specific and depends on the quantity of the polled registers. Generally, 100ms will work fine for most of the BARO Module.
Gisuportahan nga MODBUS FUNCTIONS
Table 7 Function Definitions
| Kalihokan Kodigo | Aksyon | Deskripsyon |
| 01 | Basaha ang kahimtang sa coil/port | Gibasa ang on/off status sa discrete output(s) sa ModBusSlave |
| 02 | Basaha ang kahimtang sa pag-input | Gibasa ang on/off status sa discrete input(s) sa ModBusSlave |
| 03 | Basaha ang mga naghupot sa mga rehistro | Nagbasa sa binary nga sulod sa paghupot sa rehistro (mga) sa ModBusSlave |
| 04 | Basaha ang mga rehistro sa input | Nagbasa sa binary nga mga sulod sa input register(s) sa ModBusSlave |
| 05 | Puwersa ang usa ka coil/port | Gipugos ang usa ka coil/port sa ModBusSlave sa on or off |
| 06 | Isulat ang usa ka rehistro | Nagsulat og usa ka bili ngadto sa usa ka holding register sa ModBusSlave |
| 15 | Force multiple coils/ports | Forces multiple coils/ports in the ModBusSlave to either on or off |
| 16 | Pagsulat daghang mga rehistro | Writes values into a series of holding registers in the ModBusSlave |
DATA REPRESENTATION UG REGISTER TABLES
Data values (setpoint values, parameters, sensor-specific measurement values, etc.) sent to and from the BARO Module use 16-bit and 32-bit holding (or input) registers with a 4-digit address notation. The address spaces are virtually distributed in different blocks for each data type. This is an approach to the Modbus Enron implementation. Table 8 shows the four main tables used by the BARO Module with their respective access rights. Table 9 describes the sub-blocks for each different data type representation.
Palihug timan-i nga ang ubang Modbus dataloggers migamit ug addressing gamit ang +1 offset. Kini usahay hinungdan sa kalibog ug gibase sa usa ka Modbus specification void. Kung adunay mga problema sa pagpatuman sa imong programa sa Modbus sa datalogger kanunay nga pagsulay sa pagsulay sa lainlaing mga offset sa rehistro ug mga tipo sa datos. Ang paggamit sa usa ka nahibal-an nga kantidad, sama sa temperatura, diin nahibal-an kung unsa nga kantidad ang mapaabut usa ka maayong praktis sa pagsugod sa pagsulay.
| Table 8 Modbus Primary Tables | |||
| Numero sa Pagparehistro | Matang sa lamesa | Access | Deskripsyon |
| 1xxx | Discrete Output Coils | Pagbasa/Pagsulat | on/off status or setup flags for the sensor |
| 2xxx | Discrete Input Contacts | Basaha | mga bandila sa status sa sensor |
| 3xxx | Analog Input Registers | Basaha | numerical input variables gikan sa sensor (aktuwal nga pagsukod sa sensor) |
| 4xxx | Analog Output Holding Registers | Pagbasa/Pagsulat | numerical output variables alang sa sensor (parameter, setpoint values, calibrations etc.) |
Kay example, register 3001 mao ang unang analog input register (unang data address alang sa input registers). Ang numeric value nga gitipigan dinhi usa ka 16-bit unsigned integer-type variable nga nagrepresentar sa unang sensor measurement parameter (pressure value). Ang parehas nga parameter sa pagsukod (bili sa presyur) mahimong mabasa sa rehistro 3201, apan niining higayona ingon usa ka 32-bit floating-point nga kantidad nga adunay format nga Big-Endian. Kung ang Modbus Master (Datalogger o usa ka PLC) nagsuporta lamang sa 32-bit float-values nga adunay format nga Little-Endian, nan ang usa makabasa sa parehas nga parameter sa pagsukod (sama nga kantidad sa presyur) sa rehistro 3301. Ang Virtual Sub-Blocks gituyo aron pasimplehon ang paningkamot sa user sa pagprograma sa pangutana sa Modbus sa mga sensor.
| Lamesa 9 Modbus Virtual Sub-Blocks | |||
| Numero sa Pagparehistro | Access | Gidak-on | Sub-Table Data Type |
| X001-X099 | Pagbasa/Pagsulat | 16 gamay | gipirmahan nga integer |
| X101-X199 | Pagbasa/Pagsulat | 16 gamay | unsigned integer |
| X201-X299 | Pagbasa/Pagsulat | 32 gamay | float Big-Endian format |
| X301-X399 | Pagbasa/Pagsulat | 32 gamay | float Little-Endian format |
REGISTER MAPPING
| Lamesa 10 Paghupot sa mga Rehistro | |
| 41000 (41001*) | Address sa Modbus Slave |
| Detalyadong Deskripsyon | Read or update the sensor’s modbus address |
| Type sa Data | Unsigned integer |
| Gitugotan nga Range | 1 – 247 |
| Unit | – |
| Mga komento | Ang na-update nga adres sa ulipon itago sa nonvolatile memory sa sensor |
| Lamesa 11 BARO Module Input Registers | |
| 32000 (32001*) | Soil Water Potential |
| Detalyadong Deskripsyon | Compensated tension value from tensiometer |
| Type sa Data | 32 bit floating Big-Endian |
| Gitugotan nga Range | -200 ngadto sa +200 |
| Unit | kPA |
| Mga komento | Tensiometer needs to be connected as slave |
| 32001 (32002*) | Temperatura sa Yuta |
| Detalyadong Deskripsyon | High accuracy on board temperature measurement |
| Type sa Data | 32 bit floating Big-Endian |
| Gitugotan nga Range | -30 ngadto sa +60 |
| Unit | degC |
| Mga komento | Tensiometer needs to be connected as slave |
| 32002 (32003*) | Suplay sa Sensor Voltage |
| Detalyadong Deskripsyon | On board supply voltage sukod |
| Type sa Data | 32 bit floating Big-Endian |
| Gitugotan nga Range | -10 ngadto sa +60 |
| Unit | Mga Volts |
| Mga komento | – |
| 32003 (32004*) | BARO Status |
| Detalyadong Deskripsyon | Binary status |
| Type sa Data | 32 bit floating Big-Endian |
| Gitugotan nga Range | 0/1 |
| Unit | – |
| Mga komento | – |
| 32004 (32005*) | BARO Reference Pressure |
| Detalyadong Deskripsyon | On board high accuracy barometric pressure measurement |
| Type sa Data | 32 bit floating Big-Endian |
| Gitugotan nga Range | +70 hangtod +120 |
| Unit | kPA |
| Mga komento | – |
| Talaan 11 Baro Module Input Registers (gipadayon) | |
| 32005 (32006*) | Tensiometer Pressure |
| Detalyadong Deskripsyon | Absolute pressure value from tensiometer |
| Type sa Data | 32 bit floating Big-Endian |
| Gitugotan nga Range | -200 ngadto sa +200 |
| Unit | kPA |
| Mga komento | Tensiometer needs to be connected as slave |
| 32006 (32007*) | BARO Temperatura |
| Detalyadong Deskripsyon | On board temperature measurement |
| Type sa Data | 32 bit floating Big-Endian |
| Gitugotan nga Range | -30 ngadto sa +60 |
| Unit | degC |
| Mga komento | – |
* Ang ubang mga aparato nagreport sa mga adres sa rehistro sa Modbus nga adunay offset nga +1. Tinuod kini alang sa Campbell Scientific Loggers ug Dataker loggers. Aron mabasa ang gusto nga rehistro gamita ang numero sa parenthesis.
EXAMPLE GAMIT ANG CR6 DATALOGGER UG MODBUS RTU
Ang Campbell Scientific, Inc. CR6 Measurement and Control Datalogger supports Modbus master and Modbus slave communication to integrate Modbus SCADA networks. The Modbus communications protocol facilitates the exchange of information and data between a computer/HMI software, instruments (RTUs), and Modbus-compatible sensors. The CR6 datalogger communicates exclusively in RTU mode. In a Modbus network, each slave device has a unique address. Therefore, sensor devices must be configured correctly before connecting to a Modbus Network. Addresses range from 1 to 247. Address 0 is reserved for universal broadcasts.
PAGPROGRAMA SA CR6 DATALOGGER
Ang mga Programa nga nagdagan sa CR6 (ug CR1000) Loggers gisulat sa CRBasic, usa ka pinulongan nga gimugna sa Campkampana Scientific. Kini usa ka taas nga lebel nga lengguwahe nga gidesinyo aron maghatag usa ka dali apan labi ka dali ug kusog nga pamaagi sa pagtudlo sa data logger kung giunsa ug kanus-a magkuha mga pagsukod, pagproseso sa datos, ug pagpakigsulti. Ang mga programa mahimong mugnaon gamit ang ShortCut Software o i-edit gamit ang CRBasic Editor, nga ang duha magamit alang sa pag-download isip stand-alone nga mga aplikasyon sa opisyal nga C.ampkampana Scientific weblugar (www.campbellsci.com). ShortCut Software (https://www.campbellsci.com/shortcut) CRBasic Editor (https://www.campbellsci.com/crbasiceditor)
A typical CRBasic program for a Modbus application consists of the following:
- Variables and constants declarations (public or private)
- Units declarations
- Mga parameter sa pag-configure
- Data tables declarations
- Logger Initializations
- Scan (Main Loop) with all the sensors to be quired
- Function call to the Data Tables
CR6 LOGGER RS-485 CONNECTION INTERFACE
Ang universal (U) terminal sa CR6 nagtanyag og 12 ka Channels nga nagkonektar sa halos bisan unsang matang sa sensor. Naghatag kini sa CR6 og katakus sa pagpares sa daghang mga aplikasyon ug giwagtang ang paggamit sa daghang mga eksternal nga peripheral.
Ang koneksyon sa Modbus CR6 nga gipakita sa Figure 12 naggamit sa RS-485 (A / B) nga interface nga gitaod sa mga terminal (C1-C2) ug (C3-C4). Kini nga mga interface mahimong molihok sa Half-Duplex ug Full-Duplex. Ang serial interface sa BARO Module nga gigamit alang niini nga exampAng le konektado sa (C1-C2) nga mga terminal.
BARO Module ngadto sa CR6 Datalogger Wiring Diagram
After assigning the BARO Module a unique Modbus Slave Address, it can be wired to the CR6 logger according to Figure 12. Make sure to connect the white and black wires according to their signals, respectively, to the C1 and C2 ports—the brown wire to 12V (V+) and the blue to G (GND). To control the power supply through your program, connect the brown wire directly to one of the SW12 terminals (switched 12V outputs).
EXAMPMGA PROGRAMA

SUPORTA SA KUSTOMER
Amihanang AMERIKA
Ang mga representante sa serbisyo sa kostumer magamit alang sa mga pangutana, problema, o feedback Lunes hangtod Biyernes, 7:00 am hangtod 5:00 pm oras sa Pasipiko.
- Email: support.environment@metergroup.com
- sales.environment@metergroup.com
- Telepono: +1.509.332.5600
- Fax: +1.509.332.5158
- Website: metergroup.com
EUROPE
- Customer service representatives are available for questions, problems, or feedback Monday through Friday,
- 8:00 to 17:00 Central European time.
- Email: support.europe@metergroup.com
- sales.europe@metergroup.com
- Telepono: +49 89 12 66 52 0
- Fax: + 49 89 12 66 52
- Website: metergroup.com
Kung mokontak sa METER pinaagi sa email, palihug iapil ang mosunod nga impormasyon:
- Ngalan
- Address
- Numero sa telepono
- Email adres
- Serye nga numero sa instrumento
Deskripsyon sa problema
NOTA: Para sa mga produkto nga gipalit pinaagi sa usa ka distributor, palihog kontaka ang distributor direkta alang sa tabang.
KASAYSAYAN SA REBISYON
Ang mosunod nga talaan naglista sa mga rebisyon sa dokumento.
| Rebisyon | Petsa | Compatible nga Firmware | Deskripsyon |
| 00 | 6.2025 | 1.10 | Inisyal nga pagpagawas |
FAQ
What should I do if I need a nonstandard cable length?
Contact Customer Support for assistance with nonstandard cable lengths.
How do I know which communication protocol to use for my application?
Evaluate the advantages and challenges of each protocol based on your application needs. If unsure, contact METER Customer Support for guidance.
Mga Dokumento / Mga Kapanguhaan
![]() |
METER BARO Module [pdf] Giya sa Gumagamit TEROS 31, TEROS 32, BARO Module, BARO Module, Module |
